• After each Operation, be sure to lubricate and sterilize the handpiece as soon as (within 1 hour) it's cleaned. Blood
coagulation can cause corrosion and rusting. However, do not lubricate the Micromotor. Oil could generate excessive heat
and cause damage.
• The Control Unit can be cleaned using a cloth moistened with ethanol with a concentration of 60-90%. Disconnect
the power supply before cleaning. The Control Unit and the Foot Control cannot be sterilized by any method.
• When the system is very frequently used, consider the maintenance of a small stock of replaceable parts.
• Do not disconnect the Motor Cord from the motor.
• Do not wipe or clean with high acid water. Do not immerse in any solutions.
• Follow the table right side for the motion time of motor and handpiece.
Continuous use over a long time with high torque may cause the overheating of
the handpiece, which in turn may result in an accident.
• Using the Micromotor with light 'ON' for more than 3 minutes continuously, motor
will become hot. (The maximum surface temperature might reach 45˚C by the
situation used) In such cases, stop using the Micromotor till the light part cools
down, or use the Micromotor without turning on the light.
• Applied parts for patient and/or operator is Handpiece.
• Do not look at LED light directly.
• If any abnormalities (dark, do not light on or flashing) are found for LED, stop
using it immediately and contact dealer.
• Grounding reliability can only be achieved when the equipment is connected to
an equipment receptacle marked "Hospital Only" or "Hospital Grade".
• For safety's sake, install the Control Unit in a place where the AC POWER Cord can be easily removed. (It is possible
to disconnect the control unit from the power source by removing the AC POWER Cord.)
• U.S. Federal law restricts this device to sale by or on the order of a licensed physician.
• This product listed below are delivered in a non-sterile condition and must be autoclaved prior to use. (Handpiece,
Micromotor, Handpiece Stand, Internal Spray Nozzle, Tube Holder, Nozzle Holder, Protection Plug, Calibration Bur,
Cleaning Wire)
• Turn off the Main Power Switch after each use.
• For service requirements and spare parts contact dealer.
• Use of NSK genuine pre-sterilized, disposable Irrigation Tube is recommended.
• Total recording time is about 100 minutes. (Surgic Pro + )
• Do not insert a USB flash drive during operation.
• Do not remove USB flash drive while LED lamp is on or blinking.
• Do not use data stored in USB flash drive for diagnostic purposes.
• Do not use a USB flash drive other than USB 2.0.
• This device is not guaranteed to work with all USB flash drive. It may not work with USB flash drive with security
functions or those requiring a great deal of electricity.
Use
Store (Transportation)
Between -10-50˚C (14-122˚F)
No moisture condensation in the Control Unit.
Use at outside of these limits may cause malfunction.
Temperature
Between 0-40˚C (32-104˚F)
NOTICE
Humidity
Between 10-85% RH
Between 10-85% RH
3
Operating Mode
Intermittent Operation
ON
OFF
3 min.
10 min.
Power plug below is used in North America area.
Plug Type NEMA 5-15P (Hospital Grade Type)
Atmospheric pressure
Between 700-1060 hPa
Between 500-1060 hPa