NSK Surgic Pro Serie Manuel D'utilisation page 19

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Error code
Error Mode
E0
System Error.
Excessive Current
E1
Detected.
Excessive Voltage
E2
Detected.
E3
Motor Sensor Error.
Control Unit Interior
E4
Over-heating Error.
E5
Breaking Error.
Motor Rotation
E6
Failure Error.
E7
Pump excess voltage.
Pump excess current.
E8
Foot Control
E9
abnormality.
LED excess voltage
E10
(Optic Motor).
LED excess current
E11
(Optic Motor).
Cause of Error
Memory failure.
((Surgic Pro + ))
- Erroneous memory. (Not
USB)
- USB flash drive other
than the type stipulated
(mouse, printer, etc,) is
connected.
Extended use under heavy load.
Short circuit in the motor.
Shorted cord (power line), Broken
internal circuit
Motor sensor failure (Hall IC).
Motor Cord failure.
Ingress of water into a Motor.
Overheating by extended use under
heavy load.
Operation of the Control Unit under
an extremely high temperature.
Abnormal voltage generated in the
start / stop switch circuit.
Failure in the start / stop switches
circuit.
Handpiece attachment failure.
Motor failure.
The Irrigation tube is incorrectly
fitted and loading the pump roller.
The pump fails.
Connection failure of the connector.
Failure of the inside of the Foot
Control.
Failure of LED circuit.
Check if the saline solution hasn't
penetrated the motor lamp.
Failure of LED circuit.
17
Remedy
Request repair.
Electrical contact may be insufficient.
Securely re-connect the Motor Cord and AC
Power cord.
When an error cannot be eliminated, request
repair.
Request repair.
Make sure to put a Protection Plug when
Thermo-disinfecting it.
Allow it to cool down before use.
Since heat is sufficiently radiated, periphery of
the main Control Unit should be
well-ventilated wherever possible. When an
error cannot be eliminated, request repair.
When rotation and stop are repeated in
short frequencies, a circuit may be activated
which limits acceleration at start. Wait a few
seconds and then use.
When an error cannot be eliminated, request
repair.
The chuck may be opened, or may not be
sufficiently closed. Securely close the chuck.
When an error cannot be eliminated, request
repair.
Check the Irrigation tube.
If an Error Code is displayed even when the
Irrigation tube is normal, request repair.
Check the connection state of the connector.
The Error Code is displayed even when the
connector is normally connected, request
repair.
Request repair.
Remove the saline solution completely.
Request repair.