Taking measurements using WatchBP Office (cont.)
«1st VISIT» Mode
1) Turn on the power – Turn on the device by
switching the I/O switch at the back of the device
to the ON position. and make sure Bluetooth
switch is in off position.
DC 6V 600mA
2) Apply the cuff – Properly fit one cuff to each of
the patient's arms.
Additional visual instruction can be found on the cuff
*
15
3) Switch to «1st VISIT» Mode – Slide the Mode
AUSCULT 1st FOLLOW UP
4) Adjust inflation pressure if necessary –
Switch to «1st VISIT» Mode.
AUSCULT 1st FOLLOW UP
Inflation pressure can be set manually from 140
to 280 mmHg by pressing and holding the Start/
Stop Button for 7 seconds.
Press the M Button to adjust the pressure and
press Start/Stop to save the setting.
AUSCULT 1st