PREPARING THE OIL TEST SET

1.
Expose the display and oil vessel loading areas by pulling the protective covers away. Connect the
mains powered charger to the rear socket and power it up unless the equipment is to be used
powered from its internal battery. Alternatively connect the 12 volt vehicle lead set to the rear socket
and a 12 V, negative to chassis, vehicle cigar/cigarette lighter socket only when the internal battery
is exhausted and no mains supply is available for the mains charger. Allow 15 minutes to bring the
internal battery up to a usable condition before starting tests.
2.
Switch on using the rear switch, then if necessary, press the ON/OFF front panel switch to start. The
rear switch need only be turned off when long periods of storage of over six months are expected.
The copyright screen appears on the display for a brief time and is followed by the main programme
menu screen. The selections available from the main menu are:–
OPTIONS
BS 148
BS 5730a
BS 5730a LEPLC
BS 5874
IEC 60156
IEC 156
WITHSTAND
5 MINUTE TEST
The main menu screen allocates labels to the three control keys below the display, namely (from left
to right) '▲', '▼' and 'SELECT'. The arrow head keys '▲' and '▼' allow scrolling up and down through
the menu options and the 'SELECT' key makes the highlighted option active.
Note:–
The screen shows only five menu options at any one time and those in view will depend
upon the last option used. When scrolling the highlighted option is the one ready for
selection.
3.
Select the required option and proceed as indicated by the instructions which follow on the display.
The following text explains the operation of each of the menu options. Not all of them will be of interest to
the user at any one time.
Options
The 'OPTIONS' sub-menu makes available the following:–
Scroll through and select as before.
AS 1767
ASTM D877
ASTM D1816
CEI 10-1
Γ
OCT 6581
IP 295
NFC 27-221
SABS 555
UNE 21
VDE 0370
SELF CHECK
OPTIONS
BS 148 : 1984
BS 5730a : 1979
BS 5874 : 1980
IEC 156 : 1973

Figure 4

Example Menu Screen

LANGUAGE
VIEW LAST RESULTS
DISPLAY CONTRAST
PASS/FAIL SETUP
PRINTER CONTROL
MENU
SELECT