5.4.1 Counters
The ENBIO steriliser counts the number of performed processes and uses it
to notify you about the recommended dates of replacing elements subject
to wearing down, and about required service inspections.
No.
Name
Recommended
frequency of
replacement
(cycles)
1.
HEPA filter
after 1000
By pressing the
field, you go to the counter display.
button. The service inspection value can only be reset by an
authorised service.
45
Yellow (Nearing
Red (Exceeded
the replacement
replacement
date, number
date, number of
of cycles)
cycles)
from 980
after 1000
The number of performed process-
es is on the left hand side, and on
the right hand side – the number
at which the given element should
be replaced or a service inspection
performed
. After replac-
ing a filter or seal, the values can be
reset by the user by pressing the
During regular operation, info screens concerning replacement of individu-
al elements or required service inspection are displayed alternately.
Counter values displayed in yellow or red do not prevent the device from
operation. However, exceeding the required inspection date may signifi-
cantly affect the device's operation and the load sterilisation process.
For replacing individual elements, please contact the manufacturer or
supplier.
5.5. Restarting
Restarting the device is forced when a process is aborted by the user by
pressing the
field, or when power or water supply is lost.
If the
field is selected, the following messages will be alter-
nately displayed, notifying you that the process has been aborted by the
When nearing a value when replace-
ment of an element or a service in-
spection is recommended, the values
will be highlighted in yellow.
If the limits are exceeded, the value
will be displayed in red.