or
,the current temperature of the process chamber in
the bottom left corner
the bottom right corner
played in the upper right corner of the screen
This is the expected time, which may be extended due to the mass and
type of charge.
cess, with the current stage of the
program indicated, while information
on subsequent stages is displayed in
the upper left corner of the screen.
When a program is being run, the
screen displays the temperature of
the selected sterilisation program
, the pressure currently in the chamber in
, while the process duration left is dis-
.
During the process, the
field is displayed instead of the START
field, enabling you to stop the pro-
cess at any time.
The upper left corner of the screen
displays the status of individual sub-
sequent program stages, e.g.
– Chamber lock
– Process chamber heating
If the process was completed successfully, the display will alternately show
information that the process has been completed and the load is sterile,
and that the chamber may be opened.
In the 134C FAST program, instruments are hot and wet after sterilization.
By pressing the
ATTENTION! When the process is completed, the chamber, the tray
and the load are hot. Maintain particular care and use protective
gloves to remove the load, or wait until it cools.
Performing the sterilisation process in the ENBIO device does not affect
field, you return to the start screen.
40