- Remove the test pack.
WARNING! The package will be hot.
In order to interpret the test correctly, read the instruction provided by the
test pack manufacturer.
- Open the pack and remove the chemical indicator from inside.
POSITIVE RESULT
chemical indicator has changed color on a dark uni-
form over the entire surface.
43
When the test program is completed,
the following screens are displayed
alternately.
You can safely open the steriliser's
process chamber.
When the process chamber is
opened, the start screen is displayed.
NEGATIVE RESULT
In the middle of the test field remains clear because of
the residual air in the middle of the device under test.
Any change of colour, uneven colouring of the test,
indicate the presence of air during the test cycle, caused by faulty oper-
ation of the steriliser.
Helix test
The Helix test represents sterilisation of tools with A-type holes in accord-
ance with the EN 13060 standard. It consists of a 1500 mm long tube open
on one side, and a closed test capsule on the other. The indicator strip is
located in the test capsule.
Helix test set
This test is used to assess the device's
performance in sterilising recessed
and porous charges, in particular:
• Pre-vacuum performance and the
rate and uniformity of steam pene-
tration.
• Saturated steam temperature and
pressure, reached for a specific time.