How to perform the test:
• Perform the test with an empty chamber, as per the EN 13060 standard.
• Place the test strip in the capsule. Read the instruction provided by the
test's manufacturer.
• Close the capsule.
• Place the test in the middle of the tray in the chamber.
• After completing the test, open the steriliser and remove the test.
WARNING! The package will be hot.
In order to interpret the test correctly, read the instruction provided by the
test pack manufacturer.
– Open the capsule and remove the test strip.
A part of the indicator strip did not change colour to the dark one due to
the presence of air in the capsule.
An insufficient change of colour of the indicator strip indicates the pres-
ence of air during the test cycle, caused by faulty operation of the steriliser.
POSITIVE RESULT
All fields of the indicator strip became
dark.
NEGATIVE RESULT
Part of the indicator strip did not change
color on the dark due to the presence of
air inside the capsule. Part of the indicator
strip did not change color on the dark due
to the presence of air inside the capsule.
5.4 Information menu
The information menu can be accessed by pressing the
the next service inspection can be displayed.
You can also change the date and time.
In order to set the date or time, touch the digits on the display. When a
field is selected, it starts to blink
values are displayed, up
and down
set the date and time.
Pressing the number again confirms it, and you can change another pa-
rameter.
We can choose the language in the same way by clicking on the shortcut. The
button marked with the letter B turns off and on the blue backlight in the back
of the screen Clicking the LED button launches the backlight control menu,
which is located on the sides of the device. LED lighting has two modes:
- Free mode, in which the user (by moving the sliders) sets the colors, intensity
and brightness of the light.
-Continuous mode, which indicates the stages of the entire sterilization pro-
cess with colors
LED lighting is available only in ENBIO S devices with colored housings: Mid-
night Blue, Dolly Pink, Ashy Stone, Yellow Sunflower.
field.
Here, information about the device
type, serial number, number of per-
formed processes, amount of free
memory available on the USB drive
for saving process data, and the ser-
vice menu
with the pro-
cess counters for sealing, filter and
and arrows used to change
. This way you can correctly
44