4.4 PERCENT (%) WEIGHING MODE

In normal weighing mode, press the
Press
or
SAMPLING
Put the item to be considered 100% on the pan and press the
displayed value starts flickering. Several seconds later, it stops flickering.
Take off all the weight on the pan and the display shows as below:
Then the sampling operation is finished. The balance can used for percent weighing.
Press
to weighing mode will bring back to percent weighing mode and the unit weight data is
still stored for use).
Note:
1.If a container needs to be used, press
counting mode.
2. If the unit weight is too small (less than 0.2 of a scale division), the display will show: --or-- PCS.
3.The memory of sampling size data will be cleared automatically every time the
used to select parts counting or percent weighing.

4.5 DENSITY CALCULATION MODE (Optional)

The balance can calculate sample's density through measuring sample weight in air, weight in water and water's density in
current temperature
PROCEDURE
Press the
mode. Then the display shows a temperature value.
The balance can be equipped with an optional hanging hook which can be used in
Density Calculation mode.
8
key twice to enter into percent weighing mode.
key to change to "100.00%" .
key to return to normal weighing mode. (Pressing
key 3 times in normal weighing mode to switch to density calculation
key after it returns
key to subtract its weight before entering into the
USER MANUAL BAR
key. Then the
key is
EN