Baxtran BAP Mode D'emploi page 10

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

C. SAMPLE WEIGHT IN AIR
Then the display shows "0.00g in-A" . Hang the sample on the hook (If a container is
used, press
sample's weight in air.
In-A: Indicate weight in air
The display shows as below
Press
key to confirm
D. SAMPLE WEIGHT IN WATER
Then the display shows "0.00g in-L" .Hang the sample on the hook (If a container is
used, hang the container on the hook and put it into the water, then press TARE key to
subtract the container's weight first), and then put the sample in the water. The display
will show a weight value, which indicates the sample's weight in water.
in-L: Indicate weight in water
The display shows as below
Press
key to confirm
E. Then the display will show a value which indicates the density of the sample. The
density calculation is finished. (ρ-O: density of the sample)
For example: "33.6841r-O"
Press
key to confirm and do the density calculation of next object.
Press
key to exit from the density calculation mode and return to weighing mode.
REMARKS
1. The density unit is g/cm3, so if the weighing unit is not "g" when entering into this
mode, it will automatically be changed to "g" . And when exiting from this mode, the
weighing unit will be back to the original one.
2. Press
3. Zeroing, Tarring and cancelling tare functions are available in this mode.
4. Density calculation formula:
10
key to subtract the container's weight first), the display will show the
key to exit from this mode at any time.
USER MANUAL BAR
r-O
Density of the sample
r-L
Density of the water
Sample weight in air
W
A
Sample weight in water
W
L
EN
EN

Publicité

Table des Matières
loading