EMS Tables
Emission test
RF emissions
Cispr 11
RF emissions
Cispr 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
MAG 2000 PLUS is intended for use in the electromagnetic environment specified
below. The customer or the user of MAG 2000 PLUS should assure that is used in
such environment.
Immunity test
Electrostatic
discharge (ESD)
EN 61000-4-2
Electrical
fast
transient/burst
IEC 61000-4-4
I.A.C.E.R. Srl
Electromagnetic emission
Compliance
Group 1
Class B
Class A
Complies
Complies
Electromagnetic immunity
Test
level
EN
60601-1-2
± 6kV contact
± 8kV air
± 2kV for power
supply lines
31
Electromagnetic environment –
guidance
MAG 2000 PLUS uses RF energy only
for its internal function. Therefore, its
RF emissions are very low and are not
likely to cause any interference in
nearby electronic equipment
MAG 2000 PLUS is suitable for use in
all establishments, including domestic
establishments
connected to the public low-voltage
power supply
network that supplies
buildings used for domestic purposes.
The MAG 2000 PLUS is suitable for use
in
all
establishments,
domestic establishments and those
directly connected to the public low
voltage power supply network that
supplies buildings used for domestic
purposes
Compliance level
± 6kV contact
± 8kV air
± 2kV per power
supply lines
and
those
directly
other
than
Electromagnetic
environment
–
guidance
Floors
should
be
wood, concrete or
ceramic tile. If floor
are
covered
with
synthetic
material,
the relative humidity
should be at least
30%
Mains power quality
should be at that of
a typical commercial
or
hospital
environment.
MNPG55-06