Impulses
EN 61000-4-5
Voltage dips, short
interruptions
and
voltage variations on
power supply input
lines
IEC 61000-4-11
Mains
power
electromagnetic field
EN 61000-4-8
MAG 2000 PLUS is intended for use in the electromagnetic environment specified below.
The customer or the user of MAG 2000 PLUS should assure that is used in such
environment
Immunity test
Test level EN
60601-1-2
Conducted RF
3 Veff from
EN 61000-4-6
150kHz to 80MHz
I.A.C.E.R. Srl
±1kV
differential
mode
< 5% U
T
(>95% dips of U
)
T
per 0,5 cycles
40% U
T
(60% dips of U
)
T
per 5 cycles
70% U
T
(30% dips of U
)
T
per 25 cycles
< 5% U
T
(>95% dips of U
)
T
per 5 seconds
3 A/m
r.f. immunity
Compliance
level
3 Veff from
150kHz to 80MHz
32
±1kV
differential
mode
< 5% U
T
(>95% dips of U
)
T
per 0,5 cycles
40% U
T
(60% dips of U
)
T
per 5 cycles
70% U
T
(30% dips of U
)
T
per 25 cycles
< 5% U
T
(>95% dips of U
)
T
per 5 seconds
3 A/m
Electromagnetic
environment – guidance
Portable
Mains power quality
should be at that of
a typical commercial
or
hospital
environment.
Mains power quality
should be at that of
a typical commercial
or
hospital
environment.
If the user of the
MAG2000 requires
continued operation
during power mains
interruptions,
it
is
recommended
that
MAG2000
be
powered
from
an
uninterruptible
power supply or a
battery.
Mains power quality
should be at that of
a typical commercial
or
hospital
environment.
and
mobile
RF
MNPG55-06