Medical
Prg
prg.
Yes/No
14
Yes
The indications of the electrodes' positioning are available in the Positions
manual.
TENS1 • Fast TENS (medical program)
Program also called conventional TENS, used for analgesic purposes; its
purpose is to induce the organism into blocking pain at the spine, in accordance
with the "Gate Control Theory" by Melzack and Wall. Pain impulses leave part
of the body (for example the hand) and run along the nerve tracts (through
small-diameter nerve fibres) until they reach the central nervous system where
the impulses are interpreted as pain. Conventional TENS activates large-
diameter nerve fibres, blocking the path of small-diameter nerve fibres at the
spine. Therefore, this action is mainly taken against the symptom: to simplify it
further, the wire transmitting pain information is obstructed.
Conventional TENS is a current that can be used to treat general daily pain. The
average number required to benefit from the treatment is 10/12 per day (no
contraindications in doubling this amount).
The program can be repeated at the end of the session for particularly
persistent pain. Due to the nature of the impulse the patient may experience
an addictive effect, meaning that the impulse will be felt less and less: if
necessary, the intensity can be increased by one level to counter this effect.
Session duration: 40 minutes (no less than 30/40 minutes), in a single phase.
Electrodes' positioning: form a square above the painful area as shown in
Figure1.
Intensity: to be adjusted in order to have a good solicitation of the stimulated
part, but not over the pain threshold.
IACER Srl
Description
Periarthritis
PHASE 1
PHASE 2
Total time
Total time
1 min
30 min
frequency
Frequency
150 Hz
90Hz
impulse
impulse
width
width 60µs
200µs
64
PHASE 3
Total time
10 min:
(3Hz-200µs
x 7sec
50%+ 1Hz-
200µs x 3
sec 60% +
30Hz-
200µs x 5
sec 50%) x
40 cycles
MNPG151-07