Electrodes' positioning and intensity levels
The electrodes have to be positioned to form a square over the painful zone by
using the channel 1 and 2 as shown above in Figure 1 (red or black up or down
are not important for the therapy purposes, follow the indications in the
Positions manual). The intensity should be adjusted to a level between the
thresholds of perception and pain: the maximum intensity level is the moment
in which the muscles surrounding the treated area begin to contract; over this
limit the stimulation does not become more effective, just more irritating, so it
is best to stop before that point.
Programs specifications
Medical
Prg
prg.
Yes/No
1
Yes
2
Yes
IACER Srl
Description
Conventional
Tens (fast)
Endorphinic
Tens (delayed)
Create a square area with the
electrodes over the painful zone.
Keep 4cm minimum distance
between the electrodes.
Figure 1 – Electrodes' positioning.
PHASE 1
PHASE 2
Total time
40min
frequency
90Hz
impulse
width 50µs
Total time
30min
frequency
1Hz