down, from muscles strengthening to muscle tropism recovery, from isotonic
exercises to hematomas' treatment, to treatments based on iontophoresis.
The patient population intended for electrotherapy treatment using the MIO-
IONOTENS device includes patients of both sexes, men and women, of age
(unless otherwise indicated by medical doctors). For further details, please
refer to the Contraindications section.
The CE0068 mark is only for the medical programs (see the following
paragraphs related to the detailed description of the programs).
Technical features
Characteristic
Power supply
Recharger
Isolation (EN 60601-1)
Applied part (EN 60601-1)
Protection level
Applied part to the patient
Dimensions
(length x height x depth)
Weight main body
Layout
Number output channel
Functioning
Intensity
Max output current
Impulse
Frequency
Impulses's width
Therapy
Display
Command
IACER Srl
Specifications
Rechargeable batteries AAA Ni-MH 4.8V
800mAh
power supply line AC 100-240V, 50/60Hz
200mA;
Output DC 6.8V, 300mA max.*
II
BF
IP22
Electrodes
260x176x60mm
205gr including batteries
ABS
2 independent
Continuous
adjustable
50mA, 1KΩ load each channel in REHA
programs
99mA, 1KΩ load each channel in the
remaining programs
Biphasic compensated square wave and
monophasic square wave
From 0.25 to 200Hz
From 20 to 450µs
Time depending on the program (1-90 min)
Reflective and illuminated LCD display
ABS keyboard with 9 keys
50