Cleaning / changing the piston unit
The piston unit 
from the piston frame by turning 
it with the Allen key (Fig. 11).
The piston cylinder 
up (Fig. 12) and the entire piston 
unit
 can now easily be 
20
removed from the front (Fig. 13) 
and cleaned externally using an 
alcoholic disinfectant.   R e-
lubricate the piston with original 
STARLAB piston grease (see 
section 8 Ordering Information). 
The piston unit should not be 
  d isassembled. In the event of 
  d amage, it should be fully 
replaced (see section 8 Ordering 
Information for a replacement 
piston unit).
Reassemble the pipette in the 
reverse order.