Cleaning pistons and seals
for 5ml model
Fully depress the tip ejector button 
the tip ejector 
Carefully unscrew the pipette cone 
piston unit
 in a clockwise direction. 
13
Pull the support ring
Remove the O-ring 
spring
 and the piston (Fig. 8). 
8
Note:
The piston is subject to spring tension. If
the spring or piston should become damaged,
the piston unit must be replaced (see section 8
Ordering Information for a replacement
piston unit).
Degrease the O-ring 
using an alcoholic disinfectant and re-lubri-
cate it with original STARLAB piston grease 
(see section 8   O rdering   I nformation).
The pipette filter 
pulling it out of the tip holder 
supplied with three replacement filters. 
Further replacement filters are available 
(see section 8 Ordering Information).
Reassemble the pipette in the reverse order.
Fig. 8
 down (Fig. 7).
5
 off the piston 
11
 . Now you can clean the 
14
of the piston unit 
14
is easy to remove by 
15
. The pipette is 
7
ErgoOne
 and pull 
2
 and the 
7
.
9
13
User Manual
®