Remove the O-ring 
(Fig. 5). 
9
Note:
The piston is subject to spring tension.
Degrease the piston 
disin  f ectant and re-lubricate the piston 
with original STARLAB piston grease (see 
section 8 Ordering   I nformation). 
A damaged piston must be replaced.
Reassemble the pipette in the reverse 
order.
Fig. 6
16
ErgoOne
, spring 
and piston 
7a
8
using an alcoholic 
9
Replacing seals for
models up to 1000µl
Remove the spacer ring 
100µl, 200µl and 1000µl models) 
(Fig. 5). 
Remove the sealing ring 
tip holder using the auxiliary tool 
(Fig. 6) and replace it (see section 8 
Ordering Information for replace-
ment sealing ring).
Reassemble the pipette in the 
reverse order.