Masimo Red PC Serie Mode D'emploi page 3

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 6
Red Patient Cables
Red PC Series
Red LNC Series
Reusable
INDICATIONS
The Red PC patient cables and the Red LNC patient cables have the same indications for use as the applicable sensors. Please
see the compatible sensor's directions for use for indications for use and prescribed information.
DESCRIPTION
The Red PC patient cables are used with Masimo® LNOP® sensors and the Red LNC patient cables are used with Masimo LNCS®
sensors.
The Red series of patient cables were verified using Masimo rainbow SET™ Technology.
WARNING: Masimo sensors and cables are designed for use with devices containing Masimo SET® or Masimo rainbow SET
oximetry or licensed to use Masimo sensors.
WARNINGS, CAUTIONS, AND NOTES
• Always refer to the oximeter module Operator's Manual for complete instructions or additional instructions.
• Ensure the cable is physically intact, with no broken or frayed wires or damaged parts. Visually inspect the cable and
discard if cracks or signs of deterioration are found.
• All sensors and cables are designed for use with specific monitors. Verify the compatibility of the monitor, cable and
sensor before use, otherwise degraded performance and/or patient injury can result.
• Carefully route patient cables to reduce the possibility of patient entanglement or strangulation.
• Failure to properly connect the sensor or the oximeter module to the cable will result in intermittent readings, inaccurate
results, or no reading.
• To avoid damage to the cable, always hold it by the connector rather than the cable when connecting or disconnecting
either end.
• To prevent damage, do not soak or immerse the sensor or cable in any liquid solution. Do not attempt to sterilize the
cable or sensor.
• Do not attempt to reprocess, recondition or recycle Masimo sensors or patient cables as these processes may damage
the electrical components, potentially leading to patient harm.
• Caution: Replace the cable when a replace cable message is displayed, or when a low SIQ message is consistently
displayed while monitoring consecutive patients after completing the low SIQ troubleshooting steps identified in the
monitoring device operator's manual.
• Note: The cable is provided with X-Cal® technology to minimize the risk of inaccurate readings and unanticipated loss
of patient monitoring. The cable will provide up to 17,520 hours of patient monitoring time. Replace the cable when the
patient monitoring time is exhausted. Some legacy monitoring devices do not distinguish between the cable and the
sensor. In these devices, a replace sensor or similar message could pertain either to the sensor or the cable.
INSTRUCTIONS
A. Connecting the Red patient cable to the Device
1. Connect the device connector of the patient cable to the mating connector of the device.
2. Ensure the patient cable connector is fully engaged with the device. A tactile or audible click of connection should be
heard.
B. Connecting the Red patient cable to the Sensors
Connect the Red PC Cable to the LNOP Sensor:
1. Orient the connecting tab of the sensor so that the "shiny" contacts are pointed up. Mate the logo on the sensor to the
logo on the patient cable.
2. Adhesive Sensor (Fig. 1a). Reusable Sensor (Fig. 1b). Insert the tab of the sensor into the patient cable connector until there
is a tactile or audible click of connection.
3. Gently tug on the patient cable connector to ensure a positive contact. Tape may be used to secure the cable to the patient
for ease of movement.
Connect the Red LNC Cable to the LNCS Sensor:
1. Orient the connector of the sensor to mate with the connector on the extension cable.
2. Press firmly to ensure the patient cable is fully engaged with the sensor connector (Fig. 2a).
3. Close the clear plastic cover to lock the sensor connector into place (Fig. 2b).
4. Gently tug on the patient cable connector to ensure a positive contact. Do not pull on the patient cable.
D IREC TIO N S FO R US E
Not made with natural rubber latex
LATEX
PCX-2108A
02/13
3
en
Non-Sterile
3569N-eIFU-0420

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Red lnc serie

Table des Matières