G. DIsInFectInG your PArI
reusABLe neBuLIZer
1. clean the nebulizer as described in section F . For
low-level disinfection every other treatment day,
soak all parts of the paRi Reusable Nebulizer (except
tubing and mask) in a solution of 1 part distilled white
vinegar and 3 parts hot tap water for 1 hour. You
can substitute respiratory equipment disinfectant
(like control iii
) for distilled white vinegar (follow
®
manufacturer's instructions for mixing).
2. Rinse all parts of the paRi Reusable Nebulizer
thoroughly with warm tap water and dry with a clean,
lint-free cloth. discard the vinegar solution when
disinfection is complete.
3. the paRi Reusable Nebulizer parts (except tubing and
mask) may also be disinfected by boiling them in clean
pot of water for a full 10 minutes.
4. hospitals and clinics can sterilize the paRi Reusable
Nebulizer parts (except mask and tubing) by
autoclaving (a maximum of 277° F or 136° c), follow
autoclave manufacturer's instructions.
5. dry parts on a clean, lint-free cloth.
CONTROL III
is a registered trademark of Maril Products, Inc.
®
G
H. mAIntenAnce oF your VIos
neBuLIZer system
FILter
Your vios
air filter is bright white when new. with use,
™
the filter will slowly turn grey, indicating the need for
replacement. the filter must be replaced immediately if it
gets wet or becomes clogged. a good rule to follow is to
replace your filter every 6 months – at the same time as
your paRi reusable nebulizer .
Removing: First make sure the vios
turned off. using a coin, remove the air filter cover by
turning it counter-clockwise (Fig.F). pull the cap and filter
out, then remove the filter from the cap (Fig.g-h). take
a new filter and press into cap with the open end visible.
place the cap and filter back into the front of your vios
use the coin to tighten by turning the cover clockwise.
note: Do not atteMPt to waSh anD reuSe
AiR FilteRs. Do Not use suBstitute FilteR
MateriaL coMPreSSor DaMage May reSuLt.
comPressor
use a damp cloth to wipe the exterior of your vios
housing and wing-tip
CAUTION
Never spray fluid into the ventilation slots on your vios
Fluid could cause damage to the electrical & compressor
parts and could lead to malfunction. in the event fluid
enters the unit, please contact paRi patient connection at
1.800.Fast.Neb (327.8632).
mAIntenAnce & rePAIr
Never operate your vios
incompletely assembled or damaged. use only original
paRi parts and attachments authorized by paRi for the
vios
aerosol delivery system. use of other parts and
™
attachments may damage the unit or cause improper
treatment. do not open or disassemble the compressor
unit at any time.
note: theRe ARe No seRviCeABle pARts iNsiDe
the vioS
AeRosol DeliveRY sYstem. seRviCe BY
™
anyone voiDS warranty.
H
Questions? 1.800.FAst.neB (327.8632) | PArI.com/VIos
power switch is
™
tubing.
™
if it is improperly or
™
™
.
™
™
.
™
EN-6