15.4 Configure display unit

Notes on verified weighing systems
In verified weighing systems menu item <P3CAL> will be locked.
To undo the lock, you have to break the seal and open the casing. On the circuit
board move the adjusting switch SWA1 to position „ADJ".
LOCK
SWA1
ADJ
Attention:
After destruction of the sealing the weighing system must be re-verified by an
authorised agency and a new sealing fitted before it can be reused for applications
subject to verification.
 For invoking menu item <P3CAL> see chap. 11.
 Press
to display the first menu item for setting decimal
place.
For configuration press
one.
Confirm selected menu item with
setting will be shown.
Select desired setting with
1. Decimal place, selectable 0, 0.0, 0.00, 0.000, 0.0000.
2. Scale type, selectable as single range scale, dual range
scale and multi-interval scale (See menu overview in chap.
11.2))
3. Adjustment / linearization
Adjustment or linearization is required after entering
configuration data.
For how to carry out adjustment see chap. 7.7 or
linearization see chap. 7.8
64
to select all menu items one by
and the current
and acknowledge by
.