To confirm, press
noLin = Adjustment
LineAr = Linearisation
For adjustment press
Ensure that there are no objects on the weighing pan.
Wait for stability display, then press
The currently set adjustment weight will be displayed.
Either apply the displayed adjustment weight or make changes with the help of
the navigation keys (see chap. 3.1). Confirm by
Carefully place adjustment weight in the centre of the weighing plate.
Wait for stability display, then press
After the adjustment the balance will carry out a self-test. Remove adjustment
weight during selftest, balance will return into weighing mode automatically.
KXS_KXG-TM-BA_IA-e-1932
and the current setting will be shown.
to select setting < noLin> and confirm by
.
.
.
, „LoAd" will be shown.
17