10.4 Data-Hold functions

• Menu setting:
„P4 OTH  „ANM"  „ON"", see chap. 11
 Press
+
setting will be displayed.
 Press
to select the desired setting.
hold 0
Function switched off (factory setting)
Peak value function
hold 1
This function indicates the highest load factor (peak value) of a
continuously rising load. The peak value remains in the display until it
is deleted with any key.
„Stable hold 1" mode
hold 2
The weight value is held automatically after reaching a stable value
up to the manipulation of any key in the display.
„Stable hold 2" mode
hold 3
The weight value is held in the display after reaching a stable value
until the load falls under 10d.
Animal weighing
hold 4
This function is suitable for jerky weighing procedures, see following
chapter 10.4.1
 Confirm input by
KXS_KXG-TM-BA_IA-e-1932
simultaneously when the function is enabled. The current
.