9. CONFIGURATION

Various instrument parameters can be configured. Press and hold the
Enter key for 2 seconds to go to the configuration menu.
Press the Enter key again to go to the next menu item.
The configuration menu is closed automatically following the last
menu item. It is possible to exit the configuration menu at any
time by pressing and holding the Enter key again for 2 seconds.
9.1

Adjustment (check) of measuring insert weight

This function enables adjustment of the measuring insert weight.
1. Start screen
2. A corresponding digital value is indicated for the double positi-
on of the measuring slide, as soon as movement of the measuring
slide is detected.
3. The measuring slide is considered to be balanced, whenever OK
is indicated in neutral position. The adjustment procedure can be
found in Chapter 10.3.
9.2
Measurement force configuration
The measurement force can be configured electronically from 0.75
N to 1.5 N. Press the buttons shown on the side to configure the
force. Factory setting: 1 N
9.3

Specification of number of references

A parameter setting can be made to specify the number of refe-
rences available to the user (1 ÷ 9). To this end, press the buttons
shown.
> 2 s