Contraindications:
Contraindications include using on GI endoscopes that are not Olympus or the Olympus echoendoscope that has a balloon
channel.
®
BioGuard
air/water cleaning adapter:
Use during patient procedures in place of standard air/water valve for insufflation and lens washing.
o
Prior to Use:
1.
Prior to clinical use, inspect and familiarize yourself with the device. If there is evidence of damage (i.e. deformed, cracked/broken,
damaged packaging), do not use this product, save for return, and contact your local STERIS Endoscopy Product
Specialist.
Directions for Use:
1.
Prior to clinical use, familiarize yourself with the device and read all the Instructions for Use.
BioGuard
®
suction valve:
1.
Remove the suction valve from the tray.
2.
Insert the suction valve into suction cylinder of an Olympus endoscope (see Figure 1 below).
3.
Twist valve slightly to seat in proper orientation within the cylinder.
4.
Confirm that the valve fits securely and that there is no gap between the valve end cap and the endoscope.
Note: The valve should not be able to be rotated once it is fully inserted into the cylinder.
5.
Actuate the valve by pressing and releasing a few times to confirm smooth function.
Note: Dipping in water may facilitate ease of use.
®
BioGuard
air/water valve:
1.
Remove the air/water valve from the tray.
2.
Insert the air/water valve into the air/water cylinder of an Olympus endoscope (see Figure 1 below).
3.
Confirm that the valve fits securely and that there is no gap between the valve end cap and the endoscope.
4.
Actuate the valve by pressing and releasing a few times to confirm smooth function.
Note: Dipping in water may facilitate ease of use.
®
BioShield
biopsy valve:
1.
If using the 3-piece kit (See Figure 2), attach the BioShield
2.
Securely place the biopsy valve onto the biopsy/suction channel opening of an Olympus endoscope.
Note: This biopsy valve does not accommodate a luer-lock syringe without an irrigating adaptor (part #00711131; sold
separately).
Torrent
®
irrigation scope connector:
1.
If using the 4-piece kit (See Figure 3), attach the Torrent
2.
Attach the Torrent
®
®
use of the Torrent
®
3.
Attach the Torrent
connector to the port, take care to ensure that the scope connector is aligned to the port and avoid excessive tightening.
Misalignment or excessive tightening of the scope connector could lead to the possibility of leakage.
4.
Once the scope connector is attached to the auxiliary water port of the endoscope, attach Torrent
irrigation scope connector. The Torrent
cross-contamination of the Torrent
5.
Place distal tip of endoscope into water filled receptacle.
6.
Press and hold the suction valve for a few seconds and then release.
7.
Confirm that endoscope is suctioning water.
8.
With distal tip of scope still under water, rest finger over hole in air/water valve.
9.
Confirm that air is exiting the tip of scope, which is evident by the production of air bubbles.
10. Remove distal tip of endoscope from water filled receptacle.
11. Press the air/water valve and hold for 5-10 seconds.
12. Confirm that water exits the air/water nozzle on the end of the endoscope.
®
BioGuard
air/water cleaning adapter:
1.
Remove the device from the package.
2.
Turn off the airflow regulator on the light source after initial rinsing of endoscope (per facility guidelines).
3.
Remove the air/water valve from the endoscope.
4.
Insert the BioGuard
5.
Place the distal end of the scope into a basin of water.
6.
On the light source, turn on the airflow regulator and set it to a maximum output of "high".
7.
Depress the BioGuard
instructions.
Note: This flushes the air/water channel with water.
00732820 Rev. C
irrigation scope connector to the Torrent
irrigation tubing, reference Instructions for Use.
irrigation scope connector to the auxiliary water port by screwing it onto the port. When attaching the scope
®
irrigation scope connector is manufactured with a one-way valve to minimize the risk of
®
irrigation system.
®
air/water cleaning adapter into the air/water cylinder on scope handle.
®
air/water cleaning adapter for the recommended time per the scope manufacturer's reprocessing
®
biopsy valve.
®
irrigation scope connector.
®
irrigation tubing (00711560 / 00711561 / 00711562 / 00711563). For
®
irrigation tubing to the Torrent
®
3 of 46