7. LIST OF CODED ERRORS
Error
code
Er02
Incorrect wiring or connection error: L and PE reversed
Er03
Incorrect wiring or connection error: L missing
Er04
U
voltage too high on earth electrode (danger): measurement STOPPED
F
Er05
The calculation (current measured by the probe) by (the measured voltage) is too low
Er06
The current measured by the probe is too unstable
Er07
Earth potential too high (potential danger): measurement STOPPED
Er08
Untimely interruption of the current during the measurement of Z
tripped?) - Probable cause: reversal of conductors N and PE or the installation leakage current is
too high
Er10
Volt
Volt
V
age too high on the voltage probe (danger): measurement STOPPED
Er18
Current probe not connected
Er 22
RH + RE or parasitic voltages are too high in 2P/3P measurement (the voltage exceeds the
safety limit in terms of contact voltage).
Er24
Backup memory full (action: delete stored data)
C.A 6456 earth and loop tester
Delivered in a carrying bag containing:
- 1 measuring cable for Euro mains socket,
- 1 measuring cable, 3 separate leads,
- 3 crocodile clips (red, yellow, and white),
- 3 contact pins (red, yellow, and white),
- 1 pouch with neck strap
- 1 optical communication cable
- data processing software program
a processing software program
a processing sof
- 6 LR6 batteries (1.5V)
- 1 user manual in 5 languages
C.A 6456 + T
Same as above + a earth kit comprising two rods, three cables on reels, terminated by clips
(red, 30m; blue, 30m; green, 10m) and a mallet
ACCESSORIES
- C172 current probe
- C174 current probe
- MN20 current probe
- serial printer
- Earth set (1T earth testing rod + one 30m long green lead on reel
+ a reinforced carrying case)
www.electronicaembajadores.com
Meaning
8. TO ORDER
79
or Z
(earth fault breaker
LN
LPE
P01123511
P01123513
P01120310
P01120330
P01120440
P01102903
P01101999