Compensation Of The Measuring Cables - Chauvin Arnoux C.A 6456 Notice De Fonctionnement

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Paramètre
Erase memory (totally or partially)
Type of compensation of cables
(see § 3.3)
Reference voltage for the
calculation of I
k
Value of the low current I
TEST
in "non-tripping" measurement
Threshold voltag U
L
Alarms :
Earth resistance threshold
R
E ALARM
Loop resistance or impedance
threshold
Measured current threshold
3.2.1
RESTORING THE DEFAULT CONFIGURATION
This can be used to restore the delivery configuration.
In SET- UP position:
PUSH
btn
dFLt
PUSH
btn
3.3

COMPENSATION OF THE MEASURING CABLES

IThere are 3 types of compensation for resistance of the measuring cables: "nOnE" (no compensation), "std"
(standard compensation for the cables delivered with the instrument: only the measuring cable with safety
plugs is counted), "uSEr" (user-defined compensation).
The default compensation is that of the measuring cable (standard compensation).
www.electronicaembajadores.com
Presses
MEM
MORE
x2
MORE
x3
MORE
x4
MORE
2
nd
+
MORE
2
nd
+
x2
MORE
2
nd
x3
+
MORE
TEST
TEST
for validation and programming
of the default configuration
50
Default
Values
values
see §3.6
User
Std
Std
nOne
voltage
see § 4.3.2
measured
6,9 or 12 mA
see § 4.2.2
12 mA
25 or 50 V
50 V
see § 4.2.2
see § 4.4.2
see § 4.6.2