Disinfection
Disinfect the components of your VORTEX
cleaning (only cleaned parts can be disinfected effectively).
In boiling water:
• Place the individual parts of your VORTEX
for at least 15 minutes.
Use a clean pot and fresh drinking water.
If your water is very hard, scale may be deposited on the
parts as a milky coating. This can be prevented by using
softened water.
NOTE
Plastic will melt if it comes into contact with the hot base of the
pot. Therefore, make sure there is plenty of water in the pot. This
way you will avoid damaging the components.
Using a standard thermal disinfector for baby bottles (not a
microwave oven):
Disinfection must last at least 15 minutes. Regarding disinfec-
tion and the quantity of water required for this, it is essential to
follow the instructions for use of the disinfector you are using.
Always check carefully that it is clean and fully functional.
CAUTION
A damp environment may encourage the growth of bacteria.
Therefore, remove the individual components from the pot or
disinfector as soon as disinfection has finished. Dry the parts as
described in the following. The risk of infection is reduced
when the parts are dried completely.
- 36 -
®
immediately after
®
in boiling water