Measurement range
Cuff pressure 0-300 mmHg,
Systolic 50-250 mmHg,
Diastolic 30-200 mmHg,
Pulse 40-180 beats/minute
Display accuracy
Systolic ± 3 mmHg,
Diastolic ± 3 mmHg,
Pulse ± 5 % of the value shown
Measurement inac-
Max. permissible standard deviation
curacy
according to clinical testing:
Systolic 8 mmHg /
Diastolic 8 mmHg
Memory
2 x 60 memory spaces
Dimensions
L 74 mm x W 72 mm x H 24.5 mm
Weight
Approximately 100 g (without bat-
teries)
Cuff size
135 to 195 mm
Permissible operating
+5°C to +40°C, 15-93% relative humi-
conditions
dity (non-condensing)
Permissible storage
-25°C to +70°C, ≤ 93% relative humi-
conditions
dity, 700-1050 hPa ambient pressure
Power supply
2 x 1.5V
AAA batteries
25
Battery life
For approx. 300 measurements, de-
pending on levels of blood pressure
and pump pressure
Accessories
Instructions for use, 2 x 1.5V AAA
batteries, storage box
Classification
Internal supply, IPX0, no AP or APG,
continuous operation, application part
type BF
• This device complies with European Standard EN60601-1-2
and is subject to particular precautions with regard to electro-
magnetic compatibility. Please note that portable and mobile
HF communication systems may interfere with this unit. More
details can be requested from the stated Customer Services
address or found at the end of the instructions for use.
• The device complies with the EU Medical Devices Directive
93/42/EEC, the German Medical Devices Act (Medizinpro-
duktgesetz) and the standards EN1060-1 (Non-invasive sphy-
gmomanometers – Part 1: General requirements), EN1060-3
(Non-invasive sphygmomanometers – Part 3: Supplemen-
tary requirements for electro-mechanical blood pressure
measuring systems) and IEC 80601-2-30 (Medical electrical
equipment – Part 2-30: Particular requirements for the basic
safety and essential performance of automated non-invasive
sphygmomanometers).