7 Operating instructions

CAUTIONS
The Dominant Flex is to be set up in such a way, that a separation from the mains
supply can be easily managed�
7.1
Connect Dominant
Flex to mains power
7.2
Check maximal
vacuum for functional
check
Specifications:
Altitude
above
sea level:
Max� Vacuum:
+ 2000 m
– 540 mmHg
+ 1000 m
– 630 mmHg
+ 500 m
– 668 mmHg
0 m
– 713 mmHg
(Tolerance: +/– 15 %)
18 |
Operating instructions
1�
Check the pump before use following the instruction in
chapter 6�1�
2�
3�
1�
2�
3�
– 72 kPa
– 84 kPa
– 89 kPa
– 95 kPa
2�1 Connect the power cord
to the appliance inlet at
the back of the Dominant
Flex�
2�2 Plug in the mains plug of
the power cord to a fixed
mains socket�
3�1 An internal self-test is
performed� When the
green LED lights up, the
device is ready for use�
1�1 Switch on the Dominant
Flex� The pump will
indicate a flow rate of
50 l/min�
2�1 Turn the vacuum regulator
to the right to set
maximum vacuum
3�1 Seal the end of the patient
tubing with your thumb�
3�2 Compare the maximum
vacuum according to the
specification (on left)� See
chapter 11 if the maximum
vacuum is not reached�