18 Technical specifications
high vacuum
- 95 kPa / - 713 mmHg
Tolerance: +/– 15 %
high flow
40, 50 or 60 l/min�
+/- 5 l/min�
9�3 kg
20�5 lbs
Rack version
100–240 V, 50 / 60 Hz
AC
120 W
ISO 13485
CE (93/42/EEC), IIa
H x W x D (rack version)
210 x 305 x 375 mm
8�2 x 12�0 x 14�8 inches
Measured at 0 m, atmospheric
pressure: 1013�25 hPa Please
note: vacuum levels may vary
depending on location (meters
above sea level, atmospheric
pressure and temperature)�
Transport/Storage Conditions
Operating Conditions
IP21
| 31
Technical specifications