13.5 Autoclave does not receive commands

a - AUTOCLAVE PRE-ARRANGES BAR AUTOMATIC ALIGNMENT wait for the
double beep after opening the door, then program the functions.
b - PURE WATER TANK is empty, minimum led level is on, fill with pure water
c - USED WATER TANK IS FULL. Maximum level led is on, discharge used water.

13.6 Autoclave printer does not print

a - PAPER CARRIER COVER IS NOT CLOSED CORRECTLY, on the display appears
the letter "D". Open and close the printer cover checking that is correctly closed.
b - PAPER MISSING, on the display appears the letter "P". Open the cover and insert
the roll of thermo-graphic paper.
c - ROLL OF PAPER IS MOUNTED WITH THERMO SIDE TOWARDS INSIDE, open
the cover and turn the roll of paper to the correct position.

13.7 Stains on instruments

a - INSTRUMENTS BECOME YELLOW, chemical liquid residual has with heat been
fixed on the instruments. They were not sufficiently rinsed.
b - STERILIZATION CHAMBER PRESENTS YELLOW SPOTS, instruments were
introduced in the chamber with chemical liquid which dripped and is fixed due to
heat. Rinsing was not sufficient.
c - INSTRUMENTS PRESENT WHITE SPOTS, rinsing was done with water with high
level of calcium and the instruments were not dried. For last rinse it is advisable to
used demineralized water and dry carefully the instruments.
d - INSTRUMENTS ARE DARKENED, this is caused by the fact that the instruments
have internally a strong component of carbon.