ENGLISH
14
DEMINERALIZER
14.1 Set of demineralizer system
Autoclave is already prepared also for plug IONIC SYSTEM or OSMOSI SYSTEM for
charging water into the autoclave.
These applications are accessories of autoclave.
These applications have been designed for remove minerals in the water before to charge
the autoclave.
First to plug the hydraulic connections it is necessary set the software of autoclave
following this instructions
:
Switch-on the autoclave with the MAINS (PICT.2 pos. 19.) switch and keep pressed
B-START
((during the switch on) until you see a string reporting the selected language
Press and release
B-START
to select next screnn until you setting the water load:
-CHARGING BY PUMP
-IONIC SYSTEM
-OSMOSI SYSTEM
To change this setting use B-SELECT.
To save keep pressed
B-DOOR
until you hear a "BEEP".
To exit keep pressed
B-START
until you hear a "BEEP".
WARNING
The number you can see on display, under the kind of water system used, it's a counter. It
gives you an indication about the number of cycles possible to make until the end of filter
(in case of osmosi system) or until the end of resins (in case of ionic system).
When you change the filter or the resins, it's necessary reset the number of the counter
after saving of the kind of system used.
For reset the counter it's necessary keep push the button
B-PUMP
until the acoustic signal
of the buzzer (it takes about 10 second) in the main page of the menu where you can start
the sterilization cycle.
37