Agilent Technologies TwisTorr 74FS AG Rack Mode D'emploi page 269

Table des Matières

Publicité

R1 (R2) TYPE: selects the type of reference measurement to
R1 (R2) SET VALUE: Sets the reference value at which the
R1 (R2) HISTERESYS: Sets, in percentage, the hysteresis of
R1 (R2) MASK TIME: sets the masked time (in hh:mm:ss),
R1 (R2) OUTPUT LOGIC: Selects the control logic of the
TwisTorr 74FS AG Rack Controller User Manual / 87-901-055-01
drive the relay. Possible types:
Frequency: driving frequency of the pump
Power: power supplied to the pump
Time: time from start
Normal: achievement of normal condition
Pressure: pressure read by the gauge, if present
relay R1 (R2) is activated. This parameter has no effect if
NORMAL (previous parameter) is selected as SETPOINT
TYPE.
the relay
measured from start, within which the SETPOINT R1 (R2) is
not driven even if the threshold value has been reached. This
parameter is not available if SETPOINT TYPE is set to
NORMAL
SETPOINT output.
positive logic (the relay is energized when the
measurement exceeds the reference)
negative logic (the relay is energized when the
measurement is below the reference)
Technical Information
Programming
269/300
17

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

X3508-64301X3508-64022

Table des Matières