The controller gives the error "gauge error" if the gauge connected is
not supported. The pressure can be read via display, serial line (win.
224) or REMOTE I/O on Programmable Analog Out.
If the Programmable Analog Output is configured to show the
pressure read by an external gauge, the controller doesn‟t
replicate exactly the signal coming from the gauge; it converts
the gauge signal in the range from 0 to 10V.
The conversion formula is the following:
V
out
Where:
is the gauge output voltage;
V
g
V
out
V
g_min
values can be drawn from the gauge technical specifications).
Figure 23
TwisTorr 74FS AG Rack Controller User Manual / 87-901-055-01
−
V
V
=
×
g
g
_
min
10
−
V
V
g
_
max
g
_
min
is the Analog Output voltage;
is the gauge voltage measuring range (these
V
÷
g_max
+25VDC P MAX 2W
SIGNAL
Gauge Connector
Technical Information
Programmable Analog Output Diagrams
J5
GAUGE
IDENTIFICATION
SIGNAL COMMON
COMMON
17
225/300