Agilent Technologies TwisTorr 74FS AG Rack Mode D'emploi page 264

Table des Matières

Publicité

17
Technical Information
Programming
264/300
EXTERNAL FAN CNF: Makes it possible to select driving
mode of the external cooling fan. Possible options:
ALWAYS ON: the cooling fan is always ON
OFF_LIMPOWER: 10 seconds after the pump has
reached NORMAL status, the fan is switched off and the
driving power of the pump is reduced to 50 W
SERIAL: the cooling fan is controlled directly via the
serial port.
VENT VALVE MODE: makes it possible to select driving of
the vent valve
AUTO TIME: the valve is controlled by the controller
MANUAL: the valve is controlled by the serial port.
AUTO PUMP SPEED: the valve is controlled by the
controller its king into account the pump's deceleration.
VENT VALVE DELAY: Sets the delay, expressed in seconds,
between stopping of the pump and opening of the vent valve.
VENT VALVE OPEN TIME: sets the time, expressed in
seconds, of opening of the vent valve. Enter 0 if the valve is
to be kept open indefinitely (until the next start).
LOAD GAS TYPE: selects the type of gas used in the
process. Therefore, the controller limits the power supplied
to the pump according to the setting of this parameter.
WATER COOLING: indicates whether the pump is cooled
with water or not. The controller limits the power supplied
to the pump according to this setting.
ACTIVE STOP: enables (YES) or disables (NO) active
braking.
STOP SPEED READING: enables (YES) or disables (NO) the
pump rotational frequency reading function also during the
slowing phase that follows a stop command
TwisTorr 74 FS AG Rack Controller User Manual / 87-901-055-01

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

X3508-64301X3508-64022

Table des Matières