Installation

WARNING!
The TwisTorr controller is designed for indoor use only and must be powered
with 3-wire power cord (see orderable parts table) and plug (internationally
approved) for user's safety. Use this power cord and plug in conjunction with
a properly grounded power socket to avoid electrical shock and to satisfy CE
requirements. High voltage developed in the controller can cause severe
injury or death. Before servicing the unit, disconnect the input power cable.
The TwisTorr controller can be used as a bench unit or a rack module, but it
NOTE
must be positioned so that free air can flow through the holes. Do not install or
use the controller in an environment exposed to atmospheric agents (rain, snow,
ice), dust, aggressive gases, or in explosive environments or those with a high
fire risk.
During operation, the following environmental conditions must be
respected:
See the appendix "Technical Information" for detailed Information
about the above mentioned and the other connections, and about the
options installation.
TwisTorr 84 FS AG Rack Controller User Manual / 87-901-037-01 (A)
temperature: from +5 °C to +45 °C;
relative humidity: 0 – 95 % (without condensation).
Instructions for Use

Installation

199/297