Pump Connector
A five-meter long cable is available to connect the controller to
the pump. The following figure shows the controller output
connector configuration where pins:
A-F = pump temperature sensor
B-C-D = 75 Vac 3-phase output to pump motor stator
E = EARTH PROTECTION
Figure 22
J5 – Gauge Connector
This connector is available to connect
The controller recognizes automatically the gauge connected by
reading the signal "identification" in the connector.
Following a table with the gauge supported:
Gauge model
FRG 700/702
IMG 500
PVG 500/502
FRG 720/730
IM 500
CDG 500 T1000
CDG 500 T0100
CDG 500 T0010
CDG 500 T0001
TwisTorr 84 FS AG Rack Controller User Manual / 87-901-037-01 (A)
Programmable Analog Output Diagrams
Controller-to-Pump Connector
Name showed in the display
FRGa
IMG
PVGa
FRGb
IM
CDG3
CDG2
CDG1
CDG0
Technical Information
Agilent active gauges.
17