Télécharger Imprimer la page

Agilent Technologies TwisTorr 84 FS AG Mode D'emploi page 286

Masquer les pouces Voir aussi pour TwisTorr 84 FS AG:

Publicité

17
Technical Information
Profibus Option
Diagnostic Buffer
Tab. 21
OFF SET
0
1
2
3
4
5
6
286/297
Diagnostic buffer
MEANING
B
station non exist (set by master)
0
B
station not ready (slave not ready for data exchange)
1
B
cfg fault (configuration data doesn't match)
2
B
ext diag (slave has external diagnostic data)
3
B
not supported (slave doesn't support requested function)
4
B
invalid slave response ( slave sets permanent '0')
5
B
prm fault (wrong parameter assignment)
6
B
master_lock (slave is parameterised by another master)
7
B
prm req (slave has to be re-parameterised)
0
B
stat diag (static diagnosis)
1
B
fixed '1'
2
B
wd_on (response monitoring active)
3
B
Freeze mode (received freeze command)
4
B
Synch mode (received synch command)
5
B
reserved
6
B
diag deactivated (slave is parameterised by another master)
7
Reserved
Master Add (master address after parameterisation. FF= without
parameter)
Ident number high
Ident number low
External diagnosis "Header" length indication including header
TwisTorr 84 FS AG Rack Controller User Manual / 87-901-037-01 (A)

Publicité

loading

Ce manuel est également adapté pour:

X3508-64001X3508-64002Twistorr 84 fs ag x3508-64001Twistorr 84 fs ag x3508-64002