PRESSURE UNIT: selects the unit of measurement of the
GAUGE MODEL ID: it shows the gauge model identified by the
controller.
If there are more model with the same identification the user
can select the right gauge connected
GAS TYPE : sets the gas type:
CORRECTION : if it is selected the gas type 4 (Other) you
TwisTorr 84 FS AG Rack Controller User Manual / 87-901-037-01 (A)
pressure used to show this on the display, for display on the
serial line and for setting of the setpoint and autopressure
thresholds. Possible options:
mbar
Pascal
Torr
0 : not configured
1 : Nitrogen
2 : Argon
3 : Idrogen
4 : Other
can set the correction factor of the pressure measurement
for this kind of gas. The pressure indicated by the
instrument is equal to:
P indicated = P measured x Gas correction / 10
Technical Information
Programming
269/297