- The correct operation of all valves.
- The correct operation of the reserve and the oxygen tube.
During ventilation the non-return valve could contaminate with vomit, blood or other secretions. In this
case, stop ventilation and proceed as follows:
- Rapidly compress the bag several times so that to expel the substance through strong jets of air.
If the non return valve is not clean yet:
- Dip the valve in the water and rapidly compress the bag several times with strong jets of air.
If the valve is still obstructed, throw out the bag.
Prescriptions
- Do not use the bag in presence of toxic substances.
- Remove the oxygen reserve and the supplementary valve if oxygen is not being given. The non-
removal reduces the recharge speed and consequently the ventilation frequency.
- Do not give supplementary oxygen in presence of free flames.
- Do not use oil, grease or other hydrocarbon containing substances on the bag and its components.
The supplementary oxygen, given under pressure, can cause explosions if mixed with hydrocarbons.
- The use of this device is reserved to qualified medical personnel trained in lung ventilation and cardiac
resuscitation advanced techniques.
- Before using the device on patients, make sure you are able to assemble, disassemble and use it.
- During ventilation, always check the airways pressure with a pressure gauge
Specification
Connectors:
patient port
SILICONE BAG INLET
reservoir valve
supplemental gas
inlet
Shelf Life
Dead Space:
non-rebreathing valve
adult mask
child mask
infant mask
Bag volume
Adult Model
1600ml
Child model
500ml
Infant model
280ml
Stroke volume of 1350ml can be achieved using two hands
Minimum Cycle Rate
Adult
20 breaths/min
Child
20 breaths/min
Infant
40 breaths/min
The performance characteristics for Manual Resuscitators will vary from user to user depending on a
variety of factors: ambient temperature, patient lung compliance, ventilatory frequency, size of opera-
tor's hands.
15 mm l.D.(F)/22 mm O.D.(M)
23 mm |.D.(M)
23 mm |.D.(F) (to silicone bag inlet)
25 mm O.D.(M) (to oxygen reservoir)
6 mm O.D.(M)
5 years
Pressure Relief Valve:
7ml
Child and infant
150ml
Adult
95ml
28ml
Stroke volume
700ml
300ml
150ml
Oxygen Concentration:
with reservoir
without reservoir 45% (adult and child models)
ENGLISH
40 ± 5cmH2O
60 ± 10cmH2O
Reservoir volume Suitable body weight
2500ml
2500ml
500ml
99%
90% (infant model)
> 30 kg
7-30 kg
<7 kg