ENGLISH
17
Principle Drawing (D)
GIMA Infant and Child Resuscitator are both equipped with Pressure Relief Valves, automatically pro-
vide and adjust the pressure in the lung, and keep it within 40 cmH
O +/- 5 cmH
O, any pressure exceed
2
2
this standard will cause the Pressure Relief Valve to jump off and push the pressure out to ensure the
patient's safety.
Pressure Relief Valve
Pressure Relief Valve Motion
Should higher inspiratory pressure be required the pressure relief valve may be overridden by placing
the thumb ever the valve as show as follows.
Overriding the Pressure Relief Valve
Assembly View