iNSTALLATiON
Mod. S/TS "edi" Dental Lamp
Make sure all the components listed below are included in the package received.
•
S/TS Dental Lamp (in the requested version)
•
Wrench for removing Front Screen
•
Instruction Manual
Clean the device before use (see section Cleaning the device).
This device should be installed by specialized technicians.
Power should always be cut off during installation.
LAMP ASSEMbLy, DEnTAL ChAIR vERSIon
- Install the lamp by inserting the terminal lamp pin into the specific hole on the dental chair.
Long arm 855 mm
Long arm 550 mm
LAMP wITh TRAnSFoRMER
- It should be powered directly to the mains corresponding to the characteristics
written on the data plate or the technical specifications given in the manual.
- Make sure the master switch comply with standard IEC/EN 61058-1.
LAMP wIThoUT TRAnSFoRMER
- It should be powered by alternate current at low voltage (17 V AC) using a
safety transformer compliant with EN 60601-1.
- The final medical installation should be declared compliant with IEC/EN 60601-1
by the installer.
- Make sure that upstream the lamp is installed a fuse of T 8A L 250 V. Warning:
the fuse is not supplied with the lamp.
MoUnTInG ThE LAMP on CEILInG-wALL-FLooR
- The applications are not supplied with the lamp.
CEiLiNG
40°
50°
- To mount every single application please refer to the relative instruction manual
SAFE WORKING LOAD MINIMUM BREAKING LOAD
2.92 Kg
2.56 Kg
750
L=H-1700-20
23.5 Kg.
20.5 Kg
WALL
40°
50°
21
FLOOR
750
40°
50°