CLEANiNG
The handles are not supplied sterile. Sterilize them before use. The
handles can be sterilized with saturated steam under pressure according
to the following sterilisation cycles with saturated steam under pressure
for up to maximum 200 sterilizations:
Sterilization temperature
C°
121
134
To remove sterilizable Handles proceed as follows.
Side handle open (fig 8)
Turn knob "G" counterclockwise and pull until completely unhooked. To fit, re-
verse the operation.
Forked lower handle (fig 9)
Press points "L", then pull out. To fit, reverse the operation.
Cleaning Front Screen and Parabolic Reflector
To remove Screen proceed as explained in Replacing Front Screen on page 20 fig.4.
Cleaning should be done with a soft cloth moistened with a glass-cleaning liquid,
Abrasive or detergent materials such as alcohol, trichloroethylene,
gasoline (petrol), turpentine or the like should not be used on any part
of the Lamp.
For cleaning plastic parts in the Lamp EDI Do noT USE detergents-
disinfectants containing: AMMonIUM hyDRoxIDE - SoDIUM
hyDRoxIDE - METhyLEnE CLoRIDE - METhyL ALCohoL.
Failure to comply with this indication might cause: • deterioration risk
of plastic parts with consequent breakage. In case of doubt, please contact
Customer Care Faro.
Minimum holding time
min
15
3
24
Press
kPa
207
308