Fig. B
EXPOSURE TIME
CYCLES
T4 (Minutes)
18
121°C
4
134°C
4
134°C Fast
4
134°C Flash
4
134°C Safety
20
134°C Prion
20
134°C Prion Fast
134°C Helix /
3.5
Bowie&Dick
-
Vacuum
The timing of the preheating and fractionating vacuum can vary depending on the autoclave from 25 - 35 minutes, these will add up with the timing of the cycles shown
in the table. All sterilization cycles have 3 vacuum phases except the 134 ° safety cycle that has 2 vacuum cycles.
DRYING TIME
LOAD TYPE
T5 (Minutes)
wrapped and
15
unwrapped
wrapped and
15
unwrapped
wrapped and
10
unwrapped
4
unwrapped
4
unwrapped
wrapped and
15
unwrapped
wrapped and
10
unwrapped
4
-
-
-
I WORKING
MAXIMUM LOAD
PRESSURE
(Kg)
RANGE
(bar)
7
1.04 ÷ 1.30
7
2.02 ÷ 2.38
2
2.02 ÷ 2.38
2
2.02 ÷ 2.38
7
2.02 ÷ 2.38
7
2.02 ÷ 2.38
2
2.02 ÷ 2.38
-
2.02 ÷ 2.38
-
-0.9
WORKING
TEMPERATURE
RANGE (°C)
121 ÷ 125
134 ÷ 138
134 ÷ 138
134 ÷ 138
134 ÷ 138
134 ÷ 138
134 ÷ 138
134 ÷ 138
-