Tecno-gaz ONYX B 7.0 Mode D'emploi page 85

Masquer les pouces Voir aussi pour ONYX B 7.0:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 93
15
Messages of error are emphasized through an Alfa-numerical code, consisting in a letter
and three numbers.
If an alarm message is diplayed (code "A") the cycle IS TO BE CONSIDERED
ABORTED: it will be necessary to repeat all the preparation and sterilization
operations.
To reset alarms and errors hold down at the same time the buttons under the bar indicated
with Reset.
ERROR
A 001
Cycle interrupted by the user
A 101
Vacuum not archieved in 10 min.
Vacuum not maintained on the first phase
A 111
of VACUUM TEST
Vacuum not maintained on the second
A 121
phase of VACUUM TEST
During preheating phase the autoclave did
A131
not load right water amount
A132
Error in the flow-meter operation
Pressure above the allowable limit during
A133
the water recall for the warmup phases
A 200
Error in the control of the EV operation
A 400
A 401
Error in the locking operation
A 403
A 405
A 551
Pressure out of limit
A 637
Error in the access to the memory card
MESSAGES OF ERRORS
CAUSE
ENGLISH
SOLUTION
Reset and restart the system
Check the gasket
Check the door adjustment
Reset and restart the system
Check the gasket
Check the door adjustment
Reset and restart the system
Check the gasket
Check the door adjustment
Reset and restart the system
Reset and restart the system
Reset and restart the system
Reset and restart the system
Reset and restart the system
Reset and restart the system
Let cool the room
Reset and restart the system
check that the SD card is
correctly inserted
Chech SD card: lever might
have moved on "lock" position
Reset and restart the system
39

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Ab012zxz

Table des Matières