Tecno-gaz ONYX B 7.0 Mode D'emploi page 66

Masquer les pouces Voir aussi pour ONYX B 7.0:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 93
ENGLISH
10.3 Characterisstics of the water to be used
TABLE SHOWING THE QUALITY LEVELS LAID BY THE EN13060
Evaporation residue
Silicon oxide
Iron
Cadmium
Lead
Remains of heavy metals apart from iron, cadmium,
lead
Chloride (Cl')
Phosphate (P20s)
Conductivity (at 20°C)
Ph value (acidity level)
Appearance
Hardness (and alkaline earth ions)
10.4 Loading of the materials in the autoclave
Arranging the materials to be sterilized on the provided trays, as follows:
Do not superpose the materials
Arrange the wrapped materials with the paper side facing upwards
Never bring the materials into contact with the sterilization chamber or the autoclave door
Put scissors and dental forceps with blades open
After loading the materials, close the door of the autoclave. The display will show the icon
and the message DOOR CLOSED.
20
CEN STANDARD EN13060
10
1
0.2
0.005
0.05
0.1
2
0.5
15
5
7,5
Transparent, clear, without deposits
0.02
mg/l
mg/l
mg/l
mg/l
mg/l
mg/l
mg/l
mg/l
s/cm
mmol/l