ENGLISH
Reading of the T1 probe during sterilization
A 651
above the maximum limit
Reading of the T2 probe during sterilization
A 653
above the maximum limit
A 661
Error in the probe reading
A 662
Error in the probe reading
Error for failed pressure achievement
A 701
during the first two phases of the warm-up
Error for failed pressure achievement
A 711
during the third phase of the warm-up
Reading of the T1 probe during sterilization
A 751
below the minumum limit
Reading of the T2 probe during sterilization
A 753
below the minumum limit
Temperature of the 121°C cycle out o the
A 781
maximum limit
Temperature of the 134°C cycle out o the
A 782
maximum limit
Errors of out of maximum time during the
A 801
first phases of discharge
Errors of out of maximum time during the
A 811
last phase of discharge
Cycle interrupted due to power supply
A 901
failure
It's need to make a SD CARD periodic backup.
In the event of one of the alarms persisting after some time, consult technical assistance.
40
Let cool the room
Reset and restart the system
Let cool the room
Reset and restart the system
Reset and restart the system
Reset and restart the system
Reset and restart the system
Reset and restart the system
Reset and restart the system
Reset and restart the system
Let cool the room
Reset and restart the system
Let cool the room
Reset and restart the system
Clean the filter in the room
Reset and restart the system
Clean the filter in the room
Reset and restart the system
Check the power supply
system of the machine and of
the room.
Reset and restart the system