6.9.2. Adjustment test with external weight (only PES)
During adjustment tests the balance automatically compares the saved value of the
adjustment weight with the actual value. This is only a check, i.e. no values are
changed.
Procedure:
Observe stable environmental conditions. A warming up time of ca. 1 hour is re-
quired for stabilization. Ensure that there are no objects on the weighing plate.
Operation
Display
Activate function [w�CA. 4]
w�CA�4
(see chap. 7).
a
z0
��
g