Adjusting With External Weight (Only Kern Ew And Ewb) - KERN EW Série Mode D'emploi

Masquer les pouces Voir aussi pour EW Série:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Turn the adjusting knob on the right side of the
balance to the WEIGH position.
Calibration is now finalised.
The balance will automatically return to weigh-
ing mode.

5.8.3 Adjusting with external weight (only KERN EW and EWB)

Carry out adjustment with the recommended adjusting weight (see Chapter 1 „Tech-
nical Data"). The adjustment can also be carried out with different adjusting weights
(see table), but not ideal from a metrological point of view. Information concerning
the adjusting weights is available at:
Model
EW 220-3NM
EW 420-3NM
EW 620-3NM
EW 820-2NM
EW 2200-2NM
EW 4200-2NM
EW 6200-2NM
EW 12000-1NM
EWB 220-2M
EWB 620-2M
EWB 1200-1M
EW/EG-N/EWB-e-1627
http://www.kern-sohn.com
Alternative adjusting weight
100 g
100 g
200 g
200 g
500 g
1000 g
2000 g
5000 g
100 g
200 g
500 g
CAL
buSY
End.
23

Publicité

Table des Matières
loading